CATACHEM, INC.
announces the addition of Fructosamine (glycated proteins) to the DiscretePak™
and VetSpec™ line of diagnostic chemistry reagents
CATACHEM,
INC. is pleased to announce the addition of Fructosamine (glycated proteins) to
the DiscretePak™ and VetSpec™ line of diagnostic chemistry reagents.
Fructosamine has been advocated as an alternative test for hemoglobin A1C.
Although the American Diabetes Association (ADA) recognizes both tests,
fructosamine, may be more effective in cases of sudden changes of diabetes
treatment, diabetic pregnancy and RBC loss or abnormalities.
Catachem procedure is based on, and is in excellent agreement with, the widely
accepted NBT reference method. The reagent is formulated to eliminate
interference from lipemia, uric acid, glucose, and hemolysis. This procedure is
precise, accurate and does not produce low negative values as observed with the
enzymatic procedures.
Please contact Catachem for your analyzer application and evaluation samples.
Phone: 203-335-9277
Fax: 203-335-9977
Email: catachem@catacheminc.com
URL: www.catacheminc.com
FRUCTOSAMINE
PERFORMANCE:
Sensitivity: 0.04 A-units/μmol/L
Linearity: 10-1200 μmol/L
INTERFERENCES:
Hemoglobin ≤ 100 mg/dL
Triglycerides ≤ 1200 mg/dL
Bilirubin ≤ 2.5 mg/dL
Uric Acid ≤ 5.0 mg/dL
PRECISION:
|
Fruc. |
Within-Run |
Day-Day |
Total |
|
Mean |
SD |
CV |
SD |
CV |
SD |
CV |
|
μmol |
μmol |
% |
μmol |
% |
μmol |
% |
|
215 |
9.6 |
4.4 |
15.9 |
8.0 |
15.9 |
7.7 |
|
513 |
14.2 |
2.7 |
22.3 |
4.6 |
25. |
5.2 |
|
830 |
19.3 |
2.7 |
27.1 |
3.4 |
30.8 |
3.8 |
ACCURACY:
N = 40
Range = 60.0-980 μmol/L
Mean of X = 416 μmol/L
Mean of Y = 390 μmol/L
Regression: y = 1.06 X = 8.5
Coefficient: r = 0.995
Sy.x = 28.0
|