A Monte Carlo evaluation of three Compton camera absorbers.
Summary of "A Monte Carlo evaluation of three Compton camera absorbers."
We present a quantitative study on the performance of cadmium zinc telluride (CZT), thallium-doped sodium iodide (NaI(Tl)) and germanium (Ge) detectors as potential Compton camera absorbers. The GEANT4 toolkit was used to model the performance of these materials over the nuclear medicine energy range. CZT and Ge demonstrate the highest and lowest efficiencies respectively. Although the best spatial resolution was attained for Ge, its lowest ratio of single photoelectric to multiple interactions suggests that it is most prone to inter-pixel cross-talk. In contrast, CZT, which demonstrates the least positioning error due to multiple interactions, has a comparable spatial resolution with Ge. Therefore, we modelled a Compton camera system based on silicon (Si) and CZT as the scatterer and absorber respectively. The effects of the detector parameters of our proposed system on image resolution were evaluated and our results show good agreement with previous studies. Interestingly, spatial resolution which accounted for the least image degradation at 140.5 keV became the dominant degrading factor at 511 keV, indicating that the absorber parameters play some key roles at higher energies. The results of this study have validated the predictions by An et al. which state that the use of a higher energy gamma source together with reduction of the absorber segmentation to sub-millimetre could achieve the image resolution of 5 mm required in medical imaging.
Affiliation
School of Engineering, University of Waikato, Private Bag 3105, Hamilton, 3240, New Zealand.
Journal Details
This article was published in the following journal.
Name: Australasian physical & engineering sciences in medicine / supported by the Australasian College of Physical Scientists in Medic
ISSN: 0158-9938
Pages:
Links
- PubMed Source: http://www.ncbi.nlm.nih.gov/pubmed/21710232
- DOI: http://dx.doi.org/10.1007/s13246-011-0085-1
Medical and Biotech [MESH] Definitions
Monte Carlo Method
In statistics, a technique for numerically approximating the solution of a mathematical problem by studying the distribution of some random variable, often generated by a computer. The name alludes to the randomness characteristic of the games of chance played at the gambling casinos in Monte Carlo. (From Random House Unabridged Dictionary, 2d ed, 1993)
Evaluation Studies As Topic
Studies determining the effectiveness or value of processes, personnel, and equipment, or the material on conducting such studies. For drugs and devices, CLINICAL TRIALS AS TOPIC; DRUG EVALUATION; and DRUG EVALUATION, PRECLINICAL are available.
Program Evaluation
Studies designed to assess the efficacy of programs. They may include the evaluation of cost-effectiveness, the extent to which objectives are met, or impact.
Self-evaluation Programs
Educational programs structured in such a manner that the participating professionals, physicians, or students develop an increased awareness of their performance, usually on the basis of self-evaluation questionnaires.
Nursing Evaluation Research
Research carried out by nurses that uses interviews, data collection, observation, surveys, etc., to evaluate nursing, health, clinical, and nursing education programs and curricula, and which also demonstrates the value of such evaluation.
PubMed Articles
In this paper, we study the feasibility of using the stochastic origin ensemble (SOE) algorithm for reconstructing images of secondary gammas emitted during proton radiotherapy from data measured with...
Performance evaluation of a pixellated Ge Compton camera.
An ongoing project is being carried out to develop a high purity germanium (HPGe) Compton camera for medical applications. The Compton camera offers many potential advantages over the conventional gam...
GEANT4 simulation of the effects of Doppler energy broadening in Compton imaging.
A Monte Carlo approach was used to study the effects of Doppler energy broadening on Compton camera performance. The GEANT4 simulation toolkit was used to model the radiation transport and interaction...
In this work, we investigate the use of a three-stage Compton camera to measure secondary prompt gamma rays emitted from patients treated with proton beam radiotherapy. The purpose of this study was (...
The propagation stage of uncertainty evaluation, known as the propagation of distributions, is in most cases approached by the GUM (Guide to the Expression of Uncertainty in Measurement) uncertainty f...
Clinical Trials
The utility of PEREGRINE Monte Carlo calculations for radiation treatment planning in a clinical setting will be assessed by comparing results with other fully three-dimensional programs....
Evaluation of Vancomycin Treatment of Infections Due to Staphylococcus Aureus
Objective: The objective of the study is to evaluate the ability of current vancomycin dosing strategies to attain the pharmacodynamic target of an area under the curve (AUC) to minimum i...
A Phase II Evaluation of Telesurgery Systems
This study evaluates a system of remote supervision of laparoscopic surgery. Laparoscopic surgery is performed through small holes in the abdomen called ports. A camera is passed through...
Stroke Team Remote Evaluation Using a Digital Observation Camera
The purpose of this trial is to determine if an experimental remote video camera system is an effective way for a stroke specialist to evaluate stroke patients from a distant site.
This prospective randomized study aims for determination whether focused simulator based laparoscopic camera training in novices can improve camera performance in an actual clinical situa...