Advertisement

One-Screw Fixation Provides Similar Stability to That of Two-Screw Fixation for Type II Dens Fractures.

03:01 EDT 24th May 2013 | BioPortfolio

Summary of "One-Screw Fixation Provides Similar Stability to That of Two-Screw Fixation for Type II Dens Fractures."


BACKGROUND:
Anterior screw fixation has been widely adopted for the treatment of Type II dens fractures. However, there is still controversy regarding whether one- or two-screw fixation is more appropriate. QUESTIONS/
PURPOSES:
We addressed three questions: (1) Do one- and two-screw fixation techniques differ regarding shear stiffness and rotational stiffness? (2) Can shear stiffness and rotational stiffness after screw fixation be restored to normal? (3) Does stiffness after screw fixation correlate with bone mineral density (BMD)?
METHODS:
We randomly assigned 14 fresh axes into two groups (seven axes each): one receiving one-screw fixation and another receiving two-screw fixation. Shear and torsional stiffness were measured using a nondestructive low-load test in six directions. A transverse osteotomy then was created at the base of the dens and fixed using one or two screws. Shear and torsional stiffness were tested again under the same testing conditions.
RESULTS:
Mean stiffness in all directions after screw fixation was similar in both groups. The stiffness after one- and two-screw fixation was not restored to normal: the mean shear stiffness restored ratio was less than 50% and the mean torsional stiffness restored ratio was less than 6% in both groups. BMD did not correlate with mean stiffness after screw fixation in both groups.
CONCLUSIONS:
One- and two-screw fixation for Type II dens fractures provide similar stability but neither restores normal shear or torsional stiffness. CLINICAL
RELEVANCE:
One-screw fixation might be used as an alternative to two-screw fixation. Assumed BMD should not influence surgical decision making.

Affiliation

Department of Orthopaedic Surgery, 2nd Affiliated Hospital of Zhejiang University College of Medicine, Hangzhou, Zhejiang Province, China.

Journal Details

This article was published in the following journal.

Name: Clinical orthopaedics and related research
ISSN: 1528-1132
Pages:

Links

Medical and Biotech [MESH] Definitions

Screw Worm Infection

Infection with larvae of the blow fly Cochliomyia hominivorax (Callitroga americanum), a common cause of disease in livestock in the southern and southwestern U.S.A.

Ankylosis

Fixation and immobility of a joint.

Tenodesis

Fixation of the end of a tendon to a bone, often by suturing.

Saccades

An abrupt voluntary shift in ocular fixation from one point to another, as occurs in reading.

PubMed Articles [ 7037 Associated PubMed Articles listed on BioPortfolio]

Biomechanical comparison of three techniques for fixation of tibial avulsion fractures of the anterior cruciate ligament.

PURPOSE: To evaluate the initial stability of a suture anchor fixation and to compare this with a screw fixation and pull-out suture fixation for anterior cruciate ligament tibia...

The comparative stability of screw versus plate versus screw and plate coronoid fixation.

To compare the biomechanical characteristics of screw versus plate versus both screw and plate fixation for large, type 3 O'Driscoll coronoid fractures.

Optimal double screw configuration for subtalar arthrodesis: a finite element analysis.

PURPOSE: The subtalar arthrodesis using screws has been performed to manage traumatic subtalar arthritis. Even though clinically there might not have been big difference between using single screw or...

Biomechanical effects of the extent of sacrectomy on the stability of lumbo-iliac reconstruction using iliac screw techniques: What level of sacrectomy requires the bilateral dual iliac screw technique?

BACKGROUND: Although both single and dual iliac screw techniques are used in spino-pelvic reconstruction following sacrectomy for treating sacral tumors, the basis for choosing between the two techniq...

Using a cannulated screw as a drill guide and sleeve: a simple technique for multiple-screw fixation for intracapsular femoral neck fracture.

Management of femoral neck fractures is a challenge to orthopedic surgeons. Anatomical reduction and stable fixation is mandatory for nondisplaced or impacted intracapsular femoral neck fractures. Thi...

Clinical Trials [ 819 Associated Clinical Trials listed on BioPortfolio]

A Comparison of Steel and Bioabsorbable Screw Fixation of Lisfranc Foot Injuries

I hypothesize that absorbable screw fixation of the foot's Lisfranc ligaments does not yield significant differences in postoperative foot stability, ligament function, and symptoms when c...

A Comparison of Stainless Steel and Bioabsorbable Screw Fixation of Lisfranc Foot Injuries

The Lisfranc ligaments are a group of ligaments that connect the bones of the middle portion of the foot to each other. The Lisfranc ligaments allow for a normal and stable range of motio...

Evaluation of Surgical Fixation Using Alternative Implants for the Treatment of Hip Fractures

The primary purpose of this study is to assess the impact of sliding hip screws versus cannulated screw fixation on rates of revision surgery at 2 years in individuals over 50 years of age...

Dynamic Internal Fixation of Femoral Neck Fractures

Evaluation of dynamic Plate-Screw-Systems for internal fixation after femoral neck fractures. Open Registry for Comparison of 1-screw-systems (DHS) with multi-screw-system Targon FN.

Evaluation of the PediGuard™ for Pedicle Screw Insertion

The aim of the study is to determine the relative effectiveness of the PediGuardTM , a device manufactured by SpineVision, Inc, for placement of a pedicle screw pilot (drill) hole and for...

Search BioPortfolio: