A Randomized RSA Study Comparing HXLPE to Conventional Polyethylene in Cemented Total Hip Arthroplasty

2020-03-31 04:03:51 | BioPortfolio

Summary

A prospective randomized study comparing two types of cemented acetabular components, a highly-crosslinked polyethylene or conventional polyethylene in total hip arthroplasty. Primary outcome are component wear measured by RSA, secondary outcomes are migration and complications. Patients are evaluated 5 years postoperatively.

Description

Background:

Total hip replacements is a cost effective treatment of a symptomatic degenerative hip disease. Cemented total hip replacements are the most common implants in Sweden (Swedish hip registry, 2009).

Wear of ultra high molecular weight polyethylene has been a issue with regard to implant survival. At the time of the shift of the millennium highly cross-linked polyethylene was introduced to the markets. Experimental studies had shown reduction of wear in hip simulator studies (McKellop,1999). Subsequently the intention is to reduce osteolysis and implant loosening. Single surgeon pilot studies in the 1970 (Wroblevski et al. 1996, Oonishi 1998, Grobelaar 1999) were the first clinical series with good results leading to the development of 2nd generation highly cross-linked polyethylenes which is now the almost exclusively used PE worldwide. So far medium term clinical evaluation and laboratory assessment have showed reduced wear and good mechanical properties (Digas et al. 2003, 2004, 2007, Röhrl et al. 2005, 2007, D´Antonio et al. 2005)

Production of highly crosslinked polyethylene (HXLPE) consists of several steps. First the PE material is crosslinked by being subjected to either gamma irradiation of various intensity or electron beam irradiation. The second step is to remove the free radicals created during the irradiation. This is done either by annealing (i.e. warming up the PE to just under its melting temperature) or melting (i.e. warming up the PE to above its melting temperature). Finally, the implant is sterilized either by gamma irradiation or Ethylene Oxide and packed in vacuum or inert atmosphere.There are some possible drawbacks and challenges in the process of cross-linking polyethylene with regard to preserving the mechanical properties and elimination of free radicals (Kurtz et al. 2006, Currier et al. 2007).

Submelt annealing preserves mechanical properties but on the other hand cannot eliminate all free radicals existing in the PE. Retrieval-studies of some 2nd generation highly cross-linked polyethylenes manufactured in this way have shown signs of oxidation. The in-vivo oxidation has been located in the non articulating area such as the circumference of the acetabular liner. It has been discussed that the femoral head may protect the polyethylene against oxidation, and hence prevent the possible negative consequences of remaining free radicals (Currier et al 2007).

Melting the polyethylene eliminates all free radicals but alters the mechanical properties. This has led to PE cracks and implant failure in some clinical studies (Furmanski et al. 2009), whereas it has not been seen in other.

Finally, wear particles of highly crosslinked polyethylene are more often in the submicron size compared to conventional PE. Although these particles have a higher biological activity (Fisher et al. 2006), results of recent studies show however, that the overall risk for osteolysis and aggressiveness of highly cross-linked particles is reduced (Galvin et al. 2007).

Because of differences in the manufacturing processes of highly crosslinking PE between manufacturers (i.e. different GUR used, type of irradiation, intensity of irradiation, postirradiative treatment, type of sterilization and packing, etc, etc) all highly crosslinked polyethylenes differ somewhat from each other. One can not extrapolate results of in vitro or in vivo wear studies of one manufacturer to another. Also, because of these manufacturing differences, fixation characteristics may differ between the various types of HXLPEs. Therefore it is imperative that all new highly crosslinked polyethylenes are subjected to clinical studies of their wear and fixation performance using high accuracy methods before being used extensively.

Methods:

A prospective randomized study comparing acetabular components made of HXLPE or conventional PE in total hip arthroplasty. The study will be conducted at Department of Orthopaedics, Umeå University, Umeå, Sweden.

Aim of study The aim of the study is to compare the wear and migration of a 2nd generation polyethylene from Link with conventional polyethylene in an all-poly cemented acetabular cup of the Lubinus brand using radiostereometric analysis (RSA) with 5-year follow-up.

Hypothesis The H0-hypothesis is that the difference with in wear between HXLPE and conventional PE will not increase up to 5-year follow-up.

Material Patients (age 50 years and above) with primary osteoarthritis between 2013 and 2014 at the orthopaedic department at the county hospital (Lycksele) affiliated to Umeå University hospital, Sweden. Patients were treated with a cemented total hip arthroplasty. At surgery, after the acetabulum was reamed, randomization was performed by opening of a consecutively numbered sealed envelope informing whether a cup made of HXLPE or conventional PE should be inserted. The envelopes were kept far away from the operation theatre and were opened by a research nurse contacted by telephone. Randomization was stratified for gender. Within each stratum there was block randomization with 6 permutations in each block.

Patients have been followed for 5 years postoperatively.

Implants • Acetabular cup Intervention: Link-Ticona GUR 1020 HXLPE, 32 mm inner diameter (Trade name Lubinus® X-LINKed® )

Control: Link-Chirulene GUR 1020 UHMWPE, 32 mm inner diameter (Trade name Lubinus ® UHMWPE)

• Femur stem Test and control: Lubinus® SP II®, CrCo femoral head, 32 mm diameter

Radio Sterometric Analysis (RSA) Radiostereometric analysis measurements will be performed 5 years postoperatively.

Wear will be measured as penetration of the femoral head into the polyethylene in the proximal, proximo-medial (2D pm), proximo-posterior (2D pp), and proximo-medial-posterior (3D) directions. Since the penetration during the initial 3 months represents polyethylene yield or bedding in, wear measurements will be reported from 3 months and up to 5 years postoperatively.

Quality of cup fixation will be measured as rotations of the cup around the three cardinal axes in relation to acetabulum with the postoperative examination as reference.

The RSA measurements will be analysed at Umeå University Hospital. Patient data will be saved in a specific research database at Umeå University Hospital following specific guidelines from the national data security committee.

Ethical approval The study has been approved by the Ethics Committee of Umeå University (Dnr 2011-173-31M, 2018-34-31M).

Sample size With a mean difference in wear at 2 years of 0.1 mm or more (SD ± 0.1) between the groups (p< 0.05), 25 patients in each group will give a power of > 80%. In order to cater for losses during the follow up a total of 57 patients will be enrolled.

Scientific Value All new materials should be monitored closely and introduced in small study groups using a high precision measurement technique. Many markets require such tests before acceptance for general use is given. The new highly crosslinked polyethylene from Link-Ticona is not equivalent to current highly crosslinked polyethylenes on the market, and one can not extrapolate wear results from these other designs.

Study Design

Conditions

Osteoarthritis, Hip

Intervention

Device: Randomization between prosthetic components

Location

Lycksele hospital
Lycksele
Sweden
921 37

Status

Completed

Source

Umeå University

Results (where available)

View Results

Links

Published on BioPortfolio: 2020-03-31T04:03:51-0400

Clinical Trials [1968 Associated Clinical Trials listed on BioPortfolio]

Anatomic Congruent Prosthetic Knee Design

This double-blinded randomized control study will use kinematic and kinetic measures to compare an anatomical congruent prosthetic knee design to a standard prosthetic knee design. The pat...

Comparative Effectiveness of Microprocessor Controlled and Carbon Fiber Prosthetic Feet in Transtibial Amputees

The purpose of this research is to evaluate benefits of a microprocessor controlled prosthetic ankle-foot device compared to a non-microprocessor controlled or passive carbon fiber prosthe...

Prosthetic Limbs After Leg Amputation: Alternative Method of Socket Design

People who have had a leg amputated often choose to use a prosthetic (artificial) leg. This study will evaluate a new method of making prosthetic legs for people who have had an amputatio...

Prosthetic Device in Children With Upper Congenital Limb Deficiency

This is a study being performed to evaluate the safety and effectiveness of 3D printed myoelectric prosthetic devices in children with a congenital upper limb deficiency.

Efficacy and Safety of Topical Application of Tranexamic Acid for Saving Blood Losses in Patients Subjected to Prosthetic Knee Surgery

To compare the efficacy and safety of topical tranexamic acid versus placebo in patients diagnosed with severe knee osteoarthritis who will undergo prosthetic knee surgery in terms of savi...

PubMed Articles [3326 Associated PubMed Articles listed on BioPortfolio]

Long-term follow-up of a custom-made prosthetic iris device in patients with congenital aniridia.

To investigate the effectiveness and safety of a custom-made iris prosthetic device used to reduce photic symptoms in patients with congenital aniridia.

The effect of prosthetic alignment on hip and knee joint kinetics in individuals with transfemoral amputation.

Prosthetic alignment directly affects the biomechanical loading in individuals with lower-limb amputation, and improper alignment may be contribute to the high incidence of hip and knee osteoarthritis...

Pre-operative and post-operative kinematic analysis in total knee arthroplasty. A pilot study.

Total knee replacement is the treatment of choice in knee osteoarthritis. Despite this, there is still a percentage of unsatisfied patients. Recently, prosthetic designs have been developed to improve...

Randomization Among: The Other Randomization.

A valid informed consent process for a randomized controlled trial requires the disclosure to potential participants that they will be randomized to receive the study intervention or a control interve...

Short-Term Clinical Outcomes of a Novel Corneal Prosthetic Device in a Rabbit Model.

To evaluate the clinical performance and biocompatibility of a novel synthetic corneal device.

Medical and Biotech [MESH] Definitions

Removal of an implanted therapeutic or prosthetic device.

The grafting or inserting of a prosthetic device of alloplastic material into the oral tissue beneath the mucosal or periosteal layer or within the bone. Its purpose is to provide support and retention to a partial or complete denture.

Removal of a MEDICAL DEVICE from the market due to the identification of an intrinsic property of the device that results in a serious risk to public health.

Device designed to wear externally to support muscular skeletal system in various movements such as RANGE OF MOTIONS; WEIGHT-BEARING; GAIT; and LOCOMOTION.

Removal of a MEDICAL DEVICE from the market due to a problem occurring in the manufacture or distribution of the product.

More From BioPortfolio on "A Randomized RSA Study Comparing HXLPE to Conventional Polyethylene in Cemented Total Hip Arthroplasty"

Quick Search

Relevant Topics

Joint Disorders
A joint is where two or more bones come together, like the knee, hip, elbow, or shoulder. Joints can be damaged by many types of injuries or diseases, including Arthritis - inflammation of a joint causes pain, stiffness, and swelling with ...

Orthopedics
Arthroplasty Joint Disorders Orthopedics Spinal Cord Disorders Orthopedics is the science or practice of correcting deformities caused by disease or damage to the bones and joints of the skeleton. This specialized branch of surgery may ...


Searches Linking to this Trial