Intimal Hyperplasia Evaluated by Optical Coherence Tomography (OCT) in de Novo Coronary Lesions Treated by Drug-eluting Balloon and Bare-metal Stent

2014-08-27 03:16:21 | BioPortfolio

Summary

Restenosis due to neointimal hyperplasia causes repeat target vessel revascularization in a relevant number of patients undergoing percutaneous coronary interventions (PCI). Drug-eluting stents (DES) are currently adopted to reduce the rate of restenosis; however, they may increase risk of stent thrombosis.

Experimental data and first clinical experiences showed that inhibition of neointimal hyperplasia may be obtained by local administration of anti-proliferative drugs (like paclitaxel) loaded on the surface of angioplasty balloons. Data on the efficacy of novel coronary drug-eluting balloons (DEBs) are lacking.

Aims of this open label prospective, randomized trial is to evaluate neointimal hyperplasia in patients undergoing bare-metal stent (BMS) implantation alone compared to those receiving additional DEB use and to assess if the technique of DEB use may affect the degree of neointimal hyperplasia.

Neointimal hyperplasia will be assessed by Optical coherence tomography (OCT).

Description

Background. Restenosis due to neointimal hyperplasia causes repeat target vessel revascularization in a relevant number of patients undergoing percutaneous coronary interventions (PCI). Drug-eluting stents (DES) are currently adopted to reduce the rate of restenosis and repeat revascularizations in the majority of cases. However, a known drawback that limits the clinical application of DES technology is the possible increased risk of stent thrombosis, requiring the need of a prolonged dual antiplatelet therapy. Such phenomenon is strongly related to the profound inhibition of stent strut endothelization, that leads to the presence of uncovered stent struts, and to the persistence of polymer molecules which may induce inflammatory reactions in the vessel wall.

Experimental data and first clinical experiences showed that inhibition of neointimal hyperplasia may be obtained by local administration of anti-proliferative drugs (like paclitaxel) loaded on the surface of angioplasty balloons. Accordingly, drug-eluting balloons (DEBs) are a promising tool to prevent restenosis and avoid the undesiderable persistence of polymer molecules of DES in the vessel wall, thus potentially increasing the safety of PCI Data on the efficacy of novel coronary DEBs and on the best technique to use them are lacking.

Optical coherence tomography (OCT), has a resolution of 5-10 μm, 10 times higher than intravascular ultrasound, thus allowing fine characterization of stent strut coverage and apposition and detection of minimal degree of in-stent neointima hyperplasia.

Objective of the study:

- To evaluate the degree of neointimal hyperplasia in patients undergoing bare-metal stent (BMS) implantation alone compared to those receiving additional DEB use.

- To assess if the technique of DEB use (pre-dilation or post-dilation) may affect the degree of neointimal hyperplasia after BMS implantation.

Study Design. Open label prospective randomized trial comparing the degree of neointimal hyperplasia assessed by OCT in patients treated by BMS implantation alone, BMS implantation after DEB predilation or BMS followed by DEB postdilation.

Consecutive patients undergoing BMS implantation and agreeing to enter the study will be randomized 1:1:1 to

1. BMS implantation.

2. BMS implantation after lesion predilation with DEB

3. BMS implantation followed by post-dilation with DEB. Enrolled patients will undergo a 6-month follow up coronary angiography with OCT evaluation of the stented segment. OCT analysis will be performed by an expert OCT analyst (GF) blinded to the treatment assignment.

Study population. 30 subjects.

Procedure Description. In all eligible patients, PCI with BMS implantation will be performed according to the physician standard practice. All patients will be pre-treated with dual antiplatelet therapy.

General considerations on DEB usage. DEB is mainly intended to serve as drug delivery to the vessel wall and should therefore always cover the stenotic area as well as any extended pre-treated (injured) vessel area including the target lesion and any adjacent (prox and distal) portions when these were eventually previously covered by a stent or dilated by a balloon catheter, incidentally or by intention.

DEB length and positioning within the target lesion must be therefore carefully chosen to avoid geographic miss between the DEB and such extended pre-treated vessel area.

1. BMS group procedure:

- Lesion predilation with an undersized semi-compliant balloon (balloon to artery ratio: 0.5/1).

- BMS implantation (stent to artery ration: 1.1/1).

- Post-dilation of the stented segment with a non-compliant balloon at high pressure (16-18 atm)

2. PRE-DEB group procedure:

1. Pre-dilation

- Pre-dilatation of the target lesion with an undersized semi-compliant standard PTCA balloon (balloon to artery ratio: 0.5/1)

2. DEB dilation:

- DEB diameter and pressure: nominal DEB diameter must be chosen to guarantee full vessel wall contact at a pressure close or slightly higher of the DEB nominal pressure (balloon to artery ratio: 1/1)

- DEB length: nominal DEB length must exceed 10 mm (5mm per edge) the length of the stent which is planned to be deployed

- DEB inflation time: 45 seconds

3. BMS implantation

4. Post-dilation

- Post dilatation of the stented segment must be performed with a non-compliant PTCA balloon

- Balloon diameter: nominal PTCA balloon diameter must be chosen to reach a balloon to stent ratio of 1:1 at high pressure (16-18 atm)

- Balloon length and positioning: PTCA balloon length should be shorter than the length of the deployed stent . In case of post stent edge residual stenosis post-dilation balloon must fall within are outside the stent (5mm per edge) which was the previously dilated by the DEB

3. POST-DEB group procedure:

1. Pre-dilation

- Pre-dilatation of the target lesion must be performed with an undersized semi-compliant standard PTCA balloon (balloon to artery ratio: 0.5/1)

2. BMS implantation

- stent to artery ratio: 1.1/1

- stent length must allow full coverage of the target lesion with a single stent as well as be 10 mm shorter than the DEB which the operator is planning to use next

3. Post-dilation

- Post dilatation of the stented segment must be performed with a non-compliant PTCA balloon

- Balloon diameter: nominal PTCA balloon diameter must be chosen to reach a balloon to stent ratio of 1:1 at high pressure (16-18 atm)

4. DEB-dilation

- DEB length and positioning: DEB length must be 10 mm longer than the previously deployed stent (or than the extended pre-treated area in case of former post-dilation outside the stent edges) and centred within such pre-treated length (5mm per edge)

- DEB inflation time: 45 seconds

- Balloon to stent ratio: 1.1:1 at a pressure close or slightly higher of the DEB nominal pressure

The result of the procedure will be assessed by 3-dimensional QCA.

Post-procedural management. All patients will undergo cardiac damage markers (Creatin-kinase-MB and Troponin I) assessment before the procedure, 6 hours after PCI and 24 hours after PCI. Thereafter, further blood samples will be performed only if clinically indicated.

After PCI, patients will be given aspirin (75-100 mg/die) life-long and clopidogrel (75 mg/die) for 3 months (according to the on-label prescription for DEB-treated patients).

Follow-up. Clinical follow-up will take place at 1 month (±1 week), 6 months (±2 weeks) and 1 year (±30 days). At 6-month follow-up all patients will undergo a quantitative coronary angiography (3-dimensional QCA) and Optical Coherence Tomography (OCT) study.

OCT analysis. OCT will be performed with the Imaging system M2 (LightLab Imaging Inc., Westford, Massachusetts), capable of a pullback speed of 2 mm/sec and acquisition frame rate of 15.6/sec, using a non-occlusive technique, with continuous intracoronary iso-osmolar contrast injection.

The entire stent length will be assessed and cross sectional images will be analysed every 0.5 mm.

STRUT COVERAGE The struts will be classified as uncovered if a tissue layer on the endoluminal surface is not visible or covered in the presence of visibile tissue between the endoluminal surface and the lumen.

The tissue coverage thickness will be measured in each strut as the distance from the strut endoluminal surface to the lumen. In each cross section analysed, the following parameters will be calculated: the percentage of covered struts (the number of covered struts/ total number of struts), the percentage of uncovered struts (number of uncovered struts/ total number of struts), the tissue coverage thickness (μm), the tissue coverage area (stent area - lumen area) and its percentage (tissue coverage area/stent area X 100); the tissue volume coverage (tissue coverage area x stent length) and its percentage (tissue coverage volume/stent volume X 100). To assess the pattern of coverage, the ratio between the difference of maximum and the minimum tissue thickness/maximum tissue coverage will be calculated in each frame. A ratio close to 1 indicates an asymmetric tissue coverage, on the opposite a ratio close to 0 indicates a symmetric tissue coverage.

INCOMPLETE STENT APPOSITION A single stent strut will be defined with incomplete stent apposition (ISA) when the distance between its endoluminal surface and the vessel wall will be higher than the entire strut thickness. ISA will be considered present if at least one single strut will be incompletely apposed to the vessel wall. In each OCT frame analysed, the number of struts with ISA and the maximum distance from the endoluminal stent strut to the vessel wall will be measured. The percentage of struts with ISA (number of struts with ISA/total number of struts), according to the presence/absence of tissue coverage, will also be reported. A strut will be defined as protruding when the strut will protrude into the lumen relative to the intima between the adjacent strut sections and the distance between its endoluminal surface and the vessel wall will be less than the entire strut thickness and higher than half of strut thickness. The percentage of protruding struts (number of protruding struts/total number of strut X 100) will be reported, according to the presence/absence of tissue coverage.

Sample size calculation and statistical analysis. The primary endpoint is the neointimal area. Secondary endpoints will be the percentage of uncovered struts, the percentage of struts with ISA and the percentage of protruding struts.

This OCT study is a superiority study and it is expected that additional DEB use to BMS implantation will lead to a reduction of the primary endpoint compared to BMS implantation alone.

Few information is available on neointimal proliferation after BMS implantation: two small non randomized studies reported maximal and minimal neointimal thickness (mm) at 7.3 month follow up (first study) and mean neointimal thickness at 8 month follow up (latter study) being > 4 folds higher in the BMS group compared to the rapamycin eluting stent group, although data on neointimal area are not available. A recent randomized study comparing 12 polymer coated rapamycin-eluting stents to 12 non polymer rapamycin eluting stents reported a neointimal area of 0.3 ± 0.2 mm2 in the polymer stent vs 1.2 ± 0.8 mm2in the non polymer stent, thus with a difference of 0.9 (95% CI 0.3-1.4). Based on these findings, we may hypothesize that additional DEB use will yield to a value of neointimal area close to that reported in the non polymer rapamycin eluting stent and that this would correspond to approximately 50% reduction of neointimal area in the BMS group. To detect such difference, 10 patients will be required in each group with a power of 0.9 at a two- sided type I error of 0.05. As we cannot anticipate whether the timing of additional DEB use: pre or post stenting, might have a different effect on the reduction of neointimal hyperplasia, 10 patients will be allocated in a randomized fashion to the predilatation DEB use group, 10 patients to the postdilatation DEB use group and 10 to the BMS group.

Study Design

Allocation: Randomized, Control: Active Control, Endpoint Classification: Efficacy Study, Intervention Model: Parallel Assignment, Masking: Open Label, Primary Purpose: Treatment

Conditions

Coronary Artery Disease

Intervention

drug (paclitaxel)-eluting balloon (DEB), bare-metal stent (BMS), Drug (paclitaxel)-eluting balloon (DEB)

Location

Institute of Cardiology, Catholic University of Sacred Heart
Rome
Italy
00168

Status

Recruiting

Source

Catholic University of the Sacred Heart

Results (where available)

View Results

Links

Published on BioPortfolio: 2014-08-27T03:16:21-0400

Clinical Trials [4125 Associated Clinical Trials listed on BioPortfolio]

Drug-eluting Balloon in Acute Myocardial Infarction

The investigators hypothesize that patients with Acute ST Elevation Myocardial Infarction may benefit from primary angioplasty with use of a drug eluting balloon/bare metal stent combinati...

Drug-eluting Balloon in Bifurcations Trial

The purpose of the DEBIUT study is to assess procedural, clinical and angiographic outcomes of: 1. Provisional T-stenting use for dilation the Paclitaxel-eluting PCI-balloon (DiorTM) in...

The Initial Double-Blind Drug-Eluting Stent vs Bare-Metal Stent Study.

The main objective of this study is to assess the safety and effectiveness of the sirolimus coated Bx VELOCITY stent in reducing angiographic in-stent late loss in de novo native coronary ...

Drug Eluting Stent (DES) in Primary Angioplasty

Stent implantation is the best treatment in patients with acute myocardial infarction (STEMI) referred for primary angioplasty (pPCI). However the occurrence of in stent restenosis is resp...

Efficacy Study of Drug-Eluting and Bare Metal Stents in Bypass Graft Lesions

The aim of this study is to compare the efficacy of drug-eluting stents and bare metal stents to reduce reblockage of bypass grafts after coronary stenting

PubMed Articles [13752 Associated PubMed Articles listed on BioPortfolio]

Angiographic and clinical performance of a paclitaxel coated balloon compared to a 2nd generation sirolimus eluting stent in patients with in-stent restenosis- the BIOLUX randomized controlled trial.

The optimal treatment of in-stent restenosis (ISR) remains unknown. This study evaluates the efficacy and safety of a paclitaxel coated balloon compared to a 2nd generation sirolimus eluting stent in ...

Comparison of Drug-Eluting Stent with Bare-Metal Stent Implantation in Femoropopliteal Artery Disease: A Systematic Review and Meta-Analysis.

This study aimed to perform a systematic review and meta-analysis of current evidence comparing the drug-eluting stent (DES) with the bare-metal stent (BMS) in the treatment of femoropopliteal artery ...

Comparison of drug-eluting stents and drug-coated balloon for the treatment of drug-eluting coronary stent restenosis: A randomized RESTORE trial.

This study sought to evaluate the optimal treatment for in-stent restenosis (ISR) of drug-eluting stents (DESs).

Neuro Elutax SV drug-eluting balloon versus Wingspan stent system in symptomatic intracranial high-grade stenosis: a single-center experience.

Intracranial atherosclerotic disease is a well-known cause of ischemic stroke. Following the SAMMPRIS trial, medical treatment is favored over stenting. Drug-eluting balloons (DEB) are widely used in ...

Drug-eluting stent implantation for subclavian artery reocclusion after bare-metal stent implantation: two-year outcomes.

Medical and Biotech [MESH] Definitions

Forms to which substances are incorporated to improve the delivery and the effectiveness of drugs. Drug carriers are used in drug-delivery systems such as the controlled-release technology to prolong in vivo drug actions, decrease drug metabolism, and reduce drug toxicity. Carriers are also used in designs to increase the effectiveness of drug delivery to the target sites of pharmacological actions. Liposomes, albumin microspheres, soluble synthetic polymers, DNA complexes, protein-drug conjugates, and carrier erythrocytes among others have been employed as biodegradable drug carriers.

Diminished or failed response of an organism, disease or tissue to the intended effectiveness of a chemical or drug. It should be differentiated from DRUG TOLERANCE which is the progressive diminution of the susceptibility of a human or animal to the effects of a drug, as a result of continued administration.

A component of the Executive Office of the President established by the Anti-Drug Abuse Act of 1988. The Office establishes policies, priorities, and objectives for national DRUG AND NARCOTIC CONTROL. The goals of the program are to reduce illicit drug use, manufacturing, and trafficking, drug-related crime and violence, and drug-related health consequences.

Stents that are covered with materials that are embedded with chemicals that are gradually released into the surrounding milieu.

The action of a drug that may affect the activity, metabolism, or toxicity of another drug.

More From BioPortfolio on "Intimal Hyperplasia Evaluated by Optical Coherence Tomography (OCT) in de Novo Coronary Lesions Treated by Drug-eluting Balloon and Bare-metal Stent"

Advertisement
Quick Search
Advertisement
Advertisement

 

Relevant Topics

Drug Discovery
Clinical Approvals Clinical Trials Drug Approvals Drug Delivery Drug Discovery Generics Drugs Prescription Drugs In the fields of medicine, biotechnology and pharmacology, drug discovery is the process by which drugs are dis...

Stent
A stent is a small mesh tube that's used to treat narrow or weak arteries. Arteries are blood vessels that carry blood away from your heart to other parts of your body.  A stent is placed in an artery as part of a procedure called angioplasty. &...


Searches Linking to this Trial