Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane.

06:00 EDT 3rd November 2011 | BioPortfolio

Summary of "Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane."


BACKGROUND:
The need for early identification of postural abnormalities without exposing patients to constant radiation has stimulated the development of instruments aiming to measure the spinal curvatures.
OBJECTIVE:
To verify the validity, repeatability and reproducibility of angular measures of sagittal curvatures of the spine obtained using an adapted arcometer, by comparing them with Cobb angles of the respective curvatures obtained by using X-rays.
METHODS:
52 participants were submitted to two procedures designed to evaluate the thoracic and lumbar curvatures: (1) X-ray examination from which the Cobb angles (CA) of both curvatures were obtained, and (2) measuring the angles with the arcometer (AA). Two evaluators collected the data using the arcometer, with the rods placed at T1, T12, L1 and L5 spinous processes levels in a way as to permit linear measurements which, with aid of trigonometry, supplied the AA.
RESULTS:
There was a very strong and significant correlation between AA and CA (r=0.94; p<0.01), with no-significant difference (p=0.32), for the thoracic curvature. There was a strong and significant correlation for the lumbar curvature (r=0.71; p<0.01) between AA and CA, with no-significant difference (p=0.30). There is a very strong correlation between intra-evaluator and inter-evaluator AA.
CONCLUSION:
It was possible to quantify reliably the thoracic and lumbar curvatures with the arcometer and it can thus be considered valid and reliable and for use in evaluating spinal curvatures in the sagittal plane.

Affiliation

Journal Details

This article was published in the following journal.

Name: Revista brasileira de fisioterapia (Sao Carlos (Sao Paulo, Brazil))
ISSN: 1809-9246
Pages:

Links

DeepDyve research library

PubMed Articles [7878 Associated PubMed Articles listed on BioPortfolio]

Correction of sagittal imbalance in treatment for adult degenerative scoliosis with thoracic lordosis and lumbar kyphosis: A case report.

Lumbar degenerative scoliosis (LDS) is a common spinal disease for senior citizens. However, LDS accompanied with thoracic lordosis and lumbar kyphosis (LK) is rare in clinic. No reports have reported...

Repeatability of topographic corneal thickness in keratoconus comparing Visante™ OCT and Oculus Pentacam HR(®) topographer.

To assess the repeatability and reproducibility of Visante™ OCT and Oculus Pentacam HR® in measuring central (CCT) and topographic corneal thickness (TCT) along the principle meridians in keratocon...

Echo Particle Image Velocimetry for Estimation of Carotid Artery Wall Shear Stress: Repeatability, Reproducibility and Comparison with Phase-Contrast Magnetic Resonance Imaging.

Measurement of hemodynamic wall shear stress (WSS) is important in investigating the role of WSS in the initiation and progression of atherosclerosis. Echo particle image velocimetry (echo PIV) is a n...

Repeatability and Reproducibility of Retinal Blood Flow Measurement Using a Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects.

To evaluate the repeatability and reproducibility of retinal blood flow (RBF) measurements in humans by using new auto-alignment and measurement software in a commercially available Doppler optical co...

Validation of a two-dimensional liquid chromatography method for quality control testing of pharmaceutical materials.

Despite the advantages of 2D-LC, there is currently little to no work in demonstrating the suitability of these 2D-LC methods for use in a quality control (QC) environment for good manufacturing pract...

Clinical Trials [3185 Associated Clinical Trials listed on BioPortfolio]

Evaluation of the Repeatability and Reproducibility of Corneal Epithelial Thickness Mapping With SD-OCT

Evaluate the repeatability and reproducibility of the RTVue-XR for measuring the total corneal thickness (pachymetry), the epithelial thickness, and the stromal thickness mapping in normal...

Evaluation of the Repeatability and Reproducibility of Corneal Epithelial Thickness Mapping With iVue SD-OCT

Evaluate the repeatability and reproducibility of iVue for measuring the total corneal thickness, the epithelial thickness, and the stromal thickness in normal subjects and patients with v...

Evaluation of the Repeatability and Reproducibility of AngioVue

The purpose of this study is to assess the repeatability and reproducibility of measurements of the vascular structures and the anatomical structures of the posterior pole of the eye based...

Vascular Reactivity Assessment by the AngioDefender Device -- A Precision Study

The objective of this study is to define the repeatability and reproducibility of quantifying flow-mediated dilation (FMD) of the brachial artery (BA) using the AngioDefender (AD) methodol...

Repeatability of X-Ray, MRI, and Gait Analysis

The purpose of this preliminary substudy to the parent study "Aerobic Exercise Intervention for Knee Osteoarthritis" is to determine the repeatability and reproducibility of various measur...

Medical and Biotech [MESH] Definitions

The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results.

Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output.

VERTEBRAE in the region of the lower BACK below the THORACIC VERTEBRAE and above the SACRAL VERTEBRAE.

Tapping of the subarachnoid space in the lumbar region, usually between the third and fourth lumbar vertebrae.

A noninvasive method for assessing BODY COMPOSITION. It is based on the differential absorption of X-RAYS (or GAMMA RAYS) by different tissues such as bone, fat and other soft tissues. The source of (X-ray or gamma-ray) photon beam is generated either from radioisotopes such as GADOLINIUM 153, IODINE 125, or Americanium 241 which emit GAMMA RAYS in the appropriate range; or from an X-ray tube which produces X-RAYS in the desired range. It is primarily used for quantitating BONE MINERAL CONTENT, especially for the diagnosis of OSTEOPOROSIS, and also in measuring BONE MINERALIZATION.

Quick Search
Advertisement
 
DeepDyve research library

Relevant Topic

Radiology
Radiology is the branch of medicine that studies imaging of the body; X-ray (basic, angiography, barium swallows), ultrasound, MRI, CT and PET. These imaging techniques can be used to diagnose, but also to treat a range of conditions, by allowing visuali...


Searches Linking to this Article